Analyzer for thedetermination of carbon and sulfurin organic as well as inorganic samples. The ELEMENTRAC CS-d is equipped with both an induction and a resistance furnace(ELTRA Dual Furnace Technology), covering the full range of carbon and sulfur analysis.

Four highly sensitive infrared (IR) cells allow for precise measurement of both high and low carbon and sulfur concentrations in only one analysis. The measuring range of each cell may be adapted to the user's specific requirements to ensure optimum measurement conditions for each application.

The ELEMENTRAC CS-d is supplied with the comprehensive and user-friendly ELEMENTS software which features statistics, groupings, reports, diagnosis tools and many additional functions.

FEATURES
Measured elements: carbon, sulfur
Samples: inorganic, organic
Furnace alignment: horizontal (resistance furnace) and vertical (induction furnace)
Sample carrier: ceramic boats / crucibles
Furnaces: induction furnace, above 2,000 °C; resistance furnace (ceramic tube), adjustable up to 1,550 °C (steps of 1 °C)
Detection method: solid state infrared absorption
Number of IR cells: 4
Material of IR path: gold
Typical analysis time: induction furnace 40 - 50 s;
resistance furnace 60 - 120 s
Chemicals required: magnesium perchlorate, platinized silica (alternatively copper oxide), sodium hydroxide
Gas required: compressed air (4 - 6 bar / 60 - 90 psi); oxygen 99.5 % pure (2 - 4 bar / 30 - 60 psi)
Power requirements induction furnace: 230 V, 50/60 Hz, 16 A fuse
Power requirements resistance furnace: 230 V, 50/60 Hz, 20 A fuse
Dimensions: (W x H x D) 89 x 84 x 79 cm
Weight : ~ 200 kg
Required equipment: PC, balance (resolution 0.0001g), monitor


Optional Autoloader for 36 crucibles available