Call Today 1-800-661-6700

TURBISCAN LAB

TURBISCAN LAB

Microtrac MRB |


The world reference stability analyzer TURBISCAN LAB enables fast and sensitive identification of destabilization mechanisms (creaming, sedimentation, flocculation, coalescence, and more). A temperature-controlled measurement cell allows either stability monitoring at specific storage temperatures or acceleration of the destabilization process with a temperature increase up to 60°C.

The instrument works by measuring the backscattering of light passing through a sample over time. This allows for the detection of changes in particle size, aggregation, sedimentation, or creaming. By analyzing these changes, researchers can gain insights into the stability and shelf-life of the product.

The TURBISCAN LAB can be used in both R&D and QC labs for the development and routine control of raw materials and final product formulations. The TURBISCAN technology not only detects stability evolution at an early stage (up to 1,000 times faster compared to visual testing), but also quantifies global formula alteration with the Turbiscan Stability Index (TSI) scale.

THE WORLD LEADER IN STABILITY ANALYSIS

  • Static Multiple Light Scattering (SMLS) technology
  • 1 measurement position
  • Temperature range from room temperature (RT) up to 60°C, to observe destabilizations in real usage and storage conditions
  • Accelerated destabilization detection – up to 1,000 x faster than visual observation
  • Real stability: sample stability analysis without the need for dilution or mechanical stress, in concentrations up to 95%
  • Quantitative measurements of dispersion stability and shelf-life, migration velocity, particle size diameter, and other parameters
  • One-click stability ranking with the Turbiscan Stability Index
  • Barcode reader for easy sample management
  • Convenient: Plug & Play, free license software on multiple computers, limited footprin

TECHNICAL DATA

Acquisition scan step 40 µm
Automatic sample recognition (bar-code) Yes
CE Certified Yes
Dimensions 38 x 42 x 32 cm
Standards ISO/TR 13097:2013, ISO/TR 18811:2018, ISO/TS 22107:2021, ISO/TS 21357:2022
Measured size range 10 nm - 1 mm
Measuring Principle Static Multiple Light Scattering (SMLS)
Number of Samples 1
Reproducibility / Repeatability on latex standards 0.1% / 0.05%
Sample concentration 0.0001 - 95%
Sample volume 2 - 20 mL
Software Turbisoft - Free license
Temperature range RT - 60°C
Weight 13 kg