Oil in Wastewater Analyzer - Automated OCMA-500
The OCMA-500 is a fully automated Oil content analyzer that quickly and accurately measures the concentration of oils from wastewater samples in less than 3 minutes. It requires a user to simply inject 8 ml of a water sample followed by 8 ml of S-316 solvent into a measuring cell. The analyst need only press a measure button and the instrument automatically mixes the 2 layers for a brief period after which the solvent/oil layers stratifies and the wastewater layer is diverted to a collection vessel. The solvent/oil fraction is then analyzed by NDIR and results are displayed as mg/L, mg/kg, mg/m on a 3.5’ backlit LCD.
Features:
- NDIR Analysis (non dispersive Infrared Spectroscopy)
- Measurement range: 0 mg/l to 200 mg/l
- Resolution:
0 to 99.9 mg/L - 0.1 mg/L
100 to 200 mg/L 1.0 mg/L - Repeatibility
0 to 9.9 mg/L ± 0.2 mg/L
10.0 to 99.9 mg/L ± 2 mg/L ± 1 digit
100 to 200 mg/L ± 4 mg/L ± 1 digit - Display: 3.5 inches, 320 x 240 dots, backlight color graphic LCD
- Calibration method: Zero, Span Calibration
- Extraction solvent: S-316 Solvent (non hazardous)
- Amount of extraction solvent required: 8 mL
- Operating conditions: 0- 40 ° C (non condensing)
- Power supply: AC 100V to 240 V +/- 10% 50/60 Hz
- Dimensions: 342 mm H x 200 mm W x 313 mm D
Call Today 1-800-661-6700