Oxygen / Hydrogen Analyzer ELEMENTRAC OH-p 2
Eltra | Chemical Identification
The ELEMENTRAC OH-p determines oxygen and hydrogen in inorganic samplesby inert gas fusion in an impulse furnace with temperatures in excess of 3,000 °C.
The ELEMENTRAC OH-pguarantees precise and fast sample analysis. The analyzer covers a wide range of applications such as metal,ceramics and other inorganic materials.
The ELEMENTRAC OH-pcan be supplied with up to two infrared cells with different path lengths, accommodating both high and low level oxygen analyses. Hydrogen concentrations are determined in the ELEMENTRAC OH-pby a robust and sensitive thermal conductivity cell.
Product Advantages
- Simultaneous oxygen/hydrogen determination with inert gas fusion technique
- NEW: closed gas management and optimized gas circulation for sensitive OH determination
- NEW: enlarged Schuetze Reagent tube for precise oxygen measurements
- NEW: gas flow system with electronic gas flow control and new leakage test
- NEW: water-cooled sample port system for effective removal of atmospheric gases
- flexible configurations and measuring ranges for O and H
- high sensitivity IR and TC cells with low detection limits
- short analysis time
- rapid, precise, accurate and reliable element determination
- powerful 8,5 kW impulse furnace for temperatures in excess of 3,000 °C
- economic analysis of grains without capsules
- NEW: chemicals and tubes are hidden behind a door (removable)
- NEW: powerful software supporting data and application export, with comment fields
- single and multipoint calibration (linear regression)
- NEW: cooling via tap water, heat exchanger or chiller
- New design allows operation in production control and laboratory
Features
Measured elements: Hydrogen, Oxygen
Samples: inorganic
Furnace alignment: vertical
Sample carrier: graphite crucibles
Fields of application: ceramics, engineering / electronics, steel / metallurgy
Furnace: electrode impulse furnace (max. 8,5 KW*), temperatures in excess of 3,000 °C
Detection method: solid state infrared absorption for oxygen, thermal conductivity for hydrogen
Typical analysis time: 120 - 180 s
Chemicals required: magnesium perchlorate, Schuetze reagents, sodium hydroxide
Gas required: compressed air, nitrogen 99.995 % pure, all gases with (2 - 4 bar / 30 - 60 psi)
Power requirements: 3~ 400 V, 50/60 Hz, max. 8,500 W
Dimensions (W x H x D): 57 x 77 x 63 cm
Weight: ~ 161 kg
Required equipment: balance (resolution 0.0001g), monitor, PC
Optional accessories: carrier gas purification, external chiller, gas calibration unit